The Director, Sálim Ali Centre for Ornithology and Natural History (SACON) invites quotations for the following :-

  1. PURCHASE OF DUAL RANGE ANALYTICAL BALANCE
Description Range/requirement
Maximum range 80-120g
Minimum range 20-50g
Readability <0.1mg
Repeatability for 100g <0.5mg
Repeatability for 20g <0.2mg
Linearity deviation 0.2mg
Weighing pan dimensions (mm)  80 approximate
Sensitivity offset 1.0mg (200g)
Setting time < 5s
Calibration Build in for both range

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

——————————————————————————————————————————————-

  1. PURCHASE OF DEEP FREEZER

 Type : Horizontal/Vertical

400 to 420 litre, -18o  to 20o C,

with wheels two sliding lids (solid) and storing bins

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF DIGITAL DRY BATH

LED Digital Dry Bath with Heating Block for 1.5/2.0 ml

with adapter E11336-Heating Block for 0.2, 0.5, 1.5/2.0 ml

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————

  1. PURCHASE OF REFRIGERATOR

Double door (Freezer and Refrigerator)

Capacity: 380 – 400 litre

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF SOLVENT CONCENTRATOR
Sample Position 50
Sample Volume 1 to 30ml
Temperature Range Ambient to 90 degree C with 1 degree C resolution and minimum 0.5 degree C accuracy
Time Range 1-99 minutes with 1 minute resolution
Pressure Range 0-30 psi
SS Test Tube Rack One SS test tube rack for each 30 ml, 10-12 ml, 4-7ml with 100 vials in each volume
Timer and water Temperature Micro processor controlled
Gas cut off Automatic
Disposable Nozzle tips 1000 Nos.
Tubes and Fitting for Nitrogen gas connection Required
Optional accessories Required

 

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF 20KV ONLINE UPS WITH TUBULAR BATTERIES

20 KV online UPS, PWM-IGBT technology, with three phase input  and single phase (240 DC) Output, 2 hr battery (Tubular) backup with battery stand

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF PC CONTROLLED DOUBLE BEAM UV/VIS SPECTROPHOTOMETER ON BUYBACK SCHEME

Sources: Deuterium and tungsten halogen lamps with automatic source change.

Wavelength range: Minimum 190 – 900 nm

Stray radiation: Minimum 0.02 % T at 220, 340nm

Wavelength accuracy: +/- 0.1nm increment

Wavelength setting and display: +/- 0.1nm increment

Wavelength repeatability: Minimum +/- 0.1 nm at D2 peak 656.1 nm

Bandpass: Variable bandwidth only (min. 0.5, 1.0, 2.0 nm)

Photometric repeatability: +/-0.001 Abs at 1 Abs

Photometric accuracy: ±0.004 Abs

Photometric stability: ±0.0005 Abs/ hr

Photometric noise less than ±0.0001

Resolution: Minimum 0.5 nm

Stability: Maximum 0.0002 A/h (At 500 nm, 2 s response) or better

Noise level: Maximum 0.0001Abs at 500nm RMS and slit of 1nm

Software Windows based operating Software.

Operating modes: Scan, fixed wavelength and programmable wavelength

Should provide 2 pairs of quartz cuvette

It should have the provision to analyse powder and thin film samples.

Comprehensive warranty : Minimum 3 years

Spectrophotometer available for buyback

Make : Perkin Elmer Lambda 35

Year of purchase : 2005

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF VORTEX SHAKER

Swirlex – Adjustable Vortex Shaker – E11190 with Tube adapter for 18 holes test tubes ø 10mm –E11193

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF WATER PURIFICATION SYSTEM

Should produce pure and ultra pure water separately

Ultra pure water should have the following quality;

Resitivity 18.2  Meg Ohm.cm

(@ 25 degree C )

Conductivity < 0.1uS/cm
TOC <5 ppb
Bacteria < 1cfu/mL
Pyrogen <0.01Eu/mL
RNase <0.01 ng /mL
Particulates > 0.22u <1 P/mL

Pure water should have the following quality;

Resitivity > 5 Meg Ohm.cm
Conductivity < 0.2uS/cm
TOC < 30 ppb

Should accept the feed water with TDS upto 100ppm

Should produce ultrapure water in the rate of three liters per hour

Should have conductivity meter before and after the purification system to monitor the performance and quality of the water

Deionization must be through electro deionization principle

Should have 50 liter reservoir for pure water

Should have UV treatment.

Comprehensive warranty : minimum 3 years

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF BINOCULAR LIGHT MICROSCOPE

Twin zooming objective, Total Magnification – 4x ~ 231x, Eye piece inclination 45° off vertical, Eyepieces – 10x, 15x, 20x, 33x, 45x, Optical system- Erect unreversed image, convergent angle 12. Interpupillary distance adjustable from 52 to 75mm. Eyepiece diopter seperately adjustable.

Objective0.8x ~ 3.5x (zoom ratio4.37:1), Auxillary objectives 0.5x, 0.7x, 1.5x, 2.0xEyepiece tube rotatable 360°, Available working distance – 181 mm.

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF HANDHELD FLOW METER PROBE

Sensor specifications

Velocity measurement – 0 to 10 m/s, Minimum water depth – 0.2 inches, Maximum cable length – expandable up to 15 meters, Cable material – Polyurethane jacketed

Meter specifications

Handheld with lithium ion rechargeable batteries, Internal and external memory for data storage, USB connector for data logging and up gradation, Material – Polycarbonate with a thermoplastic elastomer (TPE) overmold, Portable Carrying case for sensor and meter.

 

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF PORTABLE WATER ANALYSER

Specifications for testing multiparameters

pH (At ambient temperature): Range: -2.60 to 16.60 Accuracy: ±0.1 mV (0.01 pH units)

Resolution: 0.1 mV (0.01 pH units), ORP (At ambient temperature): Range: -1999 to +1999 mV Accuracy: ±0.5 Mv, Resolution: 0.1 mV, Conductivity (At ambient temperature): Range: 0.0 to 200 mS/cm each range Accuracy: ±0.1% FS ± digit for µS/cm to 0.1 mS/cm (range dependent), Resolution: 0.0001 mS/cm or 0.1, Dissolved Oxygen (At ambient temperature): Range: 0.00 to 90 mg/L; 0 to 550% Accuracy: ±0.2% FS (550% air saturation) ±1 digit (with 1.25 PE membrane at 10°C) Resolution: 0.01 mg/L; 0.1% air saturation Membrane Response Time (T95; 100%-0) and Flow Dependence: 1.25 mil PE = 8 seconds and 6 in/sec flow 2.0 mil PE = 17 seconds and 3 in/sec flow 1 mil Teflon® = 18 seconds and 12 in/sec flow, Temperature (At ambient temperature): Range: -10 to 100.00°C Accuracy: ±0.2% FS ±1 digit Resolution: 0.1°C Barometer: Built-in barometer

Water proof portable case, Ability to give accurate results in different environmental setups.

Calibration kit, Chargeable battery operated with long battery life, Internal Memory for on field data storage, USB cord with computer compatibility.

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF STEREOSCOPIC MICROSCOPE

Research grade fluorescence Stero microscope with a view field up to 55mm and zoom ratio of 31:1. When the 0.63X and 2X objectives are employed the zoom capability should give magnifications from 4X to 125X. To be used for brightfield, darkfield, polarization, and fluorescence imaging.

The widefield eyepieces should be adjustable to 23 degrees of inclination for user comfort and are able to be independently focused.

Transmitted light illumination –either a30W halogen bulbs or a 150W fiber optic illuminator.

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF TRINOCULAR LIGHT MICROSCOPE WITH CAMERA ATTACHMENT AND COMPUTER ACCESSIBILITY

Biological laboratory phase contrast microscope.

Objective Lenses: DIN Infinity Corrected Optical System (ICOS) U. Plan Phase 10x, 20x, s40x and s100x oil objectives mounted in smooth-operating, ball bearing mounted, quintuple nosepiece.

Viewing Head: Siedentop f-type trinocular head for camera integration. Each head to have eyetubes inclined at 30° with the left eye tube having graduated diopter settings. The interpupillary distance is adjustable between 53mm – 75mm. An 80/20 beam splitter for the trinocular tube can be engaged for photo work. (100% to eyetubes or 80% to photo tube & 20% to the eye tube.)

Eyepieces: 10x Widefield High Eyepoint eyepieces (F.N. 22 standard with a 19mm reticle mount). 15x and 20x eyepieces as an option.

Widefield High Eyepoint 10x focusable eyepiece that accept 25mm reticles.

Total Magnification: 100x, 200x, 400x, and 1000x.

Specimen Stage: preferably Ceramic coated right-handed, flat top stage to measure 171mm x 115mm and travels 78mm(X) x 52mm(Y). Ergonomically positioned coaxial drop down controls.

Illumination: 30 watt halogen Koehler illumination or LED illumination..

Light Focusing: Zernike Phase Condenser, NA 1.25 with centerable phase annuli for 10x, 20x, 40x, 100x with brightfield position and darkfiels stop for 40x objective – in quick change dovetail mount.

Servicing  and warranty : Preferable for three years

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF FTIR SPECTROMETER

Spectral Range                 7,500 to 370 cm-1 , with standard KBr beamsplitter

10,000 to 400 cm-1 , optional, with broadband KBr beamsplitter

5,000 to 200 cm-1 , optional with CsI optics

6,000 to 500 cm-1 , optional with “High Humidity” ZnSe optics

 

Resolution          Better than 1 cm-1 (apodized), optional better than 0.5 cm-1  (apodized)

Wavenumber Accuracy               Better than 0.01 cm-1 @ 2,000 cm-1

Photometric Accuracy                 Better than 0.1% T

Signal-to-Noise (Minimum)       5 Sec: >6,000:1 (= <7.2*10-5AU noise) peak-to-peak, 4 cm-1 resolution

Signal-to-Noise (Achievable)    5 Sec: >8,000:1 (= <5.4*10-5AU noise) peak-to-peak, 4 cm-1 resolution1

Min: >45,000:1 (= <9.7*10-6AU noise) peak-to-peak, 4 cm-1 resolution

Interferometer                                                Permanent aligned, High stability

Scan Speed                                         3 velocities, 2.2 – 20 kHz (1.4 – 12.7 mm/sec opd),

optional 4 velocities, 2.2 – 40 kHz (1.4 – 25.5 mm/sec opd)

Detector                                              High sensitivity DLATGS, optional LN2cooled MCT detector, easy Exchange

A/D Converter                                 24 bit

Validation                                           IVU internal validation unit

Spectrometer Size                          66.5(W) x 43.4(D) x 28.1(H) cm

Sample Compartment Size                         25.5(W) x 27(D) x 16(H) cm

Weight                                                 37 Kg

Spectrometer Power                    85 – 265 VAC, 45 – 67 Hz, 70 W

Computer Interface                       Ethernet connection

Spectroscopy Software                OPUSTM: easy to use, full GLP compliant, full 21 CFR part 11 compliant

LAST DATE FOR RECEIPT OF QUOTATIONS  : 09.01.2017


  1. PURCHASE OF STEEL DISPLAY RACK

(a)          Type – A                               Qty : 03 nos.

(b)         Type – B                              Qty : 02 nos.

LAST DATE FOR RECEIPT OF QUOTATIONS  : 16.01.2017

—————————————————————————————————————————————-

  1. PURCHASE OF FURNITURE

(a)          Revolving computer chair with cushion              Qty – 16 Nos.

(b)         Laboratory Stool                                                             Qty – 06 Nos.

(c)          Wooden Stage for class room                                   Qty – One

Size : 14.5’ x 7’  Height : ½’

Quality : 18mm plywood double for bottom support and 12mm plywood for top with

Quality terracotta laminated sheet

LAST DATE FOR RECEIPT OF QUOTATIONS  : AT THE EARLIEST